3.2f Assembling the Oxygen Equipment
3.2f Assembling the Oxygen Equipment
Assembling oxygen equipment
When the equipment has been prepared, it must be assembled as follows:
Check the cylinder valve is off firmly.
Use a properly fitted regulator with the correct pin ‘indexed’ yolk for B and C size cylinders.
Check the ‘O’ ring (Bodok seal) and ‘valve seat’ are clean and undamaged.
Carefully fit the yoke over the valve and align the pins on the regulator with the valve body.
Firmly tighten the regulator-locking screw against the valve bod
Attach the oxygen therapy or mechanical resuscitator devices to the oxygen outlet.
Ensure the flow metre is off or set to ‘zero’.
Turn the cylinder valve on slowly.
Turn the oxygen flow on and check the system for leaks and correct function.
Test the system by checking that oxygen is flowing into the mask or cannula.
Check that the bag mask or demand valve resuscitator is operating correctly.
Continue with applying oxygen to the patient in accordance with your training.
If carrying out a check only, turn the flow meter and the cylinder valve off.
Discard single use items, such as nasal cannula, after use.
Disinfect or sterilise any re-usable items that have been used on a patient.
Do not empty a cylinder completely
as it will allow moisture to develop inside.