3.2f Assembling the Oxygen Equipment

Assembling oxygen equipment

When the equipment has been prepared, it must be assembled as follows:
  • Check the cylinder valve is off firmly.
  • Use a properly fitted regulator with the correct pin ‘indexed’ yolk for B and C size cylinders.
  • Check the ‘O’ ring (Bodok seal) and ‘valve seat’ are clean and undamaged.
  • Carefully fit the yoke over the valve and align the pins on the regulator with the valve body.
  • Firmly tighten the regulator-locking screw against the valve bod
  • Attach the oxygen therapy or mechanical resuscitator devices to the oxygen outlet.
  • Ensure the flow metre is off or set to ‘zero’.
  • Turn the cylinder valve on slowly.
  • Turn the oxygen flow on and check the system for leaks and correct function.
  • Test the system by checking that oxygen is flowing into the mask or cannula.
  • Check that the bag mask or demand valve resuscitator is operating correctly.
  • Continue with applying oxygen to the patient in accordance with your training.
  • If carrying out a check only, turn the flow meter and the cylinder valve off.
  • Discard single use items, such as nasal cannula, after use.
  • Disinfect or sterilise any re-usable items that have been used on a patient.
    • Do not empty a cylinder completely
    as it will allow moisture to develop inside.